Dual energy CT

Changed by Domenico De Santis, 12 Feb 2019

Updates to Article Attributes

Title was changed:
Dual energy-energy CT
Body was changed:

Dual energy-energy CT utilises(DECT) utilises two separate energy sets to examine the different attenuation properties of matter, having a significant advantage over traditional single energy CT. Independent attenuation values at two energy sets can create virtual non-contrast images from contrast-enhanced imaging as well as delineate the composition of renal calculi and arterial plaque 13.

There are three different dual energyDECT technologies available:

  • dual source dual energy Dual-source DECT
    • twoTwo x-ray tubes producing different voltages offset at approximately 90 degrees°
  • single source dual energySingle-source helicalDECT
    • a singleTwo spiral scans are consecutively acquired at different tube potentials.
  • Single-source twin-beam DECT
    • A two-material filter splits the x-ray beam into high-energy and low-energy spectra on the z-axis before it reaches the patient.
  • Single-source sequential DECT
    • Each x-ray tube with fast switching voltage otherwise known as kVp switchingrotation is performed at high- and low- tube potential.
  • single source dualSingle-source rapid switching DECT
    • The x-ray tube switches between high- and low- tube potential multiple times within the same rotation.
  • Dual-layer DECT
    • A "sandwich" detector absorbs high-energy photons with its top layer, and low-energy photons with its bottom layer
Dual energy-energy

X-ray photons primarily interact with matter via the photoelectric effect, and Compton scattering producing the diagnostic images used in medicine today.

When an atom undergoes the photoelectric effect, the electron from that respected K-shell otherwise referred to as the inner shell is ejected via the incident photon. As that electron is excited, vacant space is 'filled' by a neighbouring electron, releasing energy as a photoelectron.

In short, when a photon has sufficient energy to overcome the electron's binding energy in the K-shell, that atom undergoes the photoelectric effect.

Each substance owns a unique K-shell binding energy; known as the K-edge. There is a significant spike in attenuation that results just beyond the energy of the K-edge, this peak is unique to every material and holds valuable information about the substance's composition.

The different photoelectric energies and K-edges are the bread and butter of dual-energy CT. Although most elements in the human body have very low K-edges (0.01-0.53 keV), elements like iodine and calcium have  K-edges of 33.2 keV and 4.0 keV respectively, making them sufficiently larger than surrounding structures and are particularly important in the clinical setting 1-3.  

For instance, at 80 kVp a structure that contains no (introduced) iodine, such as the liver, has an attenuation based on its K-edge of x, yet when iodine (33.2 keV) is introduced into that same structure, it has a higher attenuation of y bringing it closer to 80 kVp.

As 80 kVp is closer to 33.2 keV than 140 kVp, the structures containing iodine will retain less attenuation as the kVp progressed beyond the K-edge of iodine. Therefore, when using two energies, it is possible to delineate structures based solely on their attenuation differences between 80 kVp and 140 kVp.

A dual x-ray source, tube A (140 kVp) and tube B (80 kVp or 100 kVp) with an angular offset of 90 degrees are preferred offsets for a dual source scanner in the current literature 1-5.

See also 

  • -<p><strong>Dual energy CT </strong>utilises two separate energy sets to examine the different attenuation properties of matter, having a significant advantage over traditional single energy CT. Independent attenuation values at two energy sets can create virtual non-contrast images from contrast-enhanced imaging as well as delineate the composition of renal calculi and arterial plaque <sup>13</sup>.</p><p>There are three different dual energy technologies available:</p><ul>
  • -<li>dual source dual energy <ul><li>two x-ray tubes producing different voltages offset at 90 degrees</li></ul>
  • +<p><strong>Dual-energy CT </strong>(DECT) utilises two separate energy sets to examine the different attenuation properties of matter, having a significant advantage over traditional single energy CT. Independent attenuation values at two energy sets can create virtual non-contrast images from contrast-enhanced imaging as well as delineate the composition of renal calculi and arterial plaque <sup>13</sup>.</p><p>There are different DECT technologies available:</p><ul>
  • +<li>
  • +<em>Dual-source DECT</em><ul><li>Two x-ray tubes producing different voltages offset at approximately 90°</li></ul>
  • -<li>single source dual energy<ul><li>a single x-ray tube with fast switching voltage otherwise known as kVp switching</li></ul>
  • +<li>Single-source <em>helical</em> <em>DECT</em><ul><li>Two spiral scans are consecutively acquired at different tube potentials.</li></ul>
  • -<li>single source dual layer</li>
  • -</ul><h5>Dual energy</h5><p>X-ray photons primarily interact with matter via the <a href="/articles/photoelectric-effect">photoelectric effect</a>, and <a href="/articles/compton-effect">Compton</a> scattering producing the diagnostic images used in medicine today.</p><p>When an atom undergoes the photoelectric effect, the electron from that respected K-shell otherwise referred to as the inner shell is ejected via the incident photon. As that electron is excited, vacant space is 'filled' by a neighbouring electron, releasing energy as a photoelectron.</p><p>In short, when a photon has sufficient energy to overcome the electron's binding energy in the K-shell, that atom undergoes the photoelectric effect.</p><p>Each substance owns a unique K-shell binding energy; known as the K-edge. There is a significant spike in attenuation that results just beyond the energy of the K-edge, this peak is unique to every material and holds valuable information about the substance's composition.</p><p>The different photoelectric energies and K-edges are the bread and butter of dual-energy CT. Although most elements in the human body have very low K-edges (0.01-0.53 keV), elements like iodine and calcium have  K-edges of 33.2 keV and 4.0 keV respectively, making them sufficiently larger than surrounding structures and are particularly important in the clinical setting <sup>1-3</sup>.  </p><p>For instance, at 80 kVp a structure that contains no (introduced) iodine, such as the liver, has an attenuation based on its K-edge of x, yet when iodine (33.2 keV) is introduced into that same structure, it has a higher attenuation of y bringing it closer to 80 kVp.</p><p>As 80 kVp is closer to 33.2 keV than 140 kVp, the structures containing iodine will retain less attenuation as the kVp progressed beyond the K-edge of iodine. Therefore, when using two energies, it is possible to delineate structures based solely on their attenuation differences between 80 kVp and 140 kVp.</p><p>A dual x-ray source, tube A (140 kVp) and tube B (80 kVp or 100 kVp) with an angular offset of 90 degrees are preferred offsets for a dual source scanner in the current literature <sup>1-5</sup>.</p><h4>See also </h4><ul><li><a href="/articles/dual-energy-ct-clinical-applications-1">clinical applications of dual energy CT</a></li></ul>
  • +<li>Single-source <em>twin-beam DECT</em><ul><li>A two-material filter splits the x-ray beam into high-energy and low-energy spectra on the z-axis before it reaches the patient.</li></ul>
  • +</li>
  • +<li>Single-source <em>sequential</em> DECT<ul><li>Each x-ray tube rotation is performed at high- and low- tube potential.</li></ul>
  • +</li>
  • +<li>Single-source <em>rapid switching DECT</em><ul><li>The x-ray tube switches between high- and low- tube potential multiple times within the same rotation.</li></ul>
  • +</li>
  • +<li>
  • +<em>Dual-layer DECT</em><ul><li>A "sandwich" detector absorbs high-energy photons with its top layer, and low-energy photons with its bottom layer</li></ul>
  • +</li>
  • +</ul><h5>Dual-energy</h5><p>X-ray photons primarily interact with matter via the <a href="/articles/photoelectric-effect">photoelectric effect</a>, and <a href="/articles/compton-effect">Compton</a> scattering producing the diagnostic images used in medicine today.</p><p>When an atom undergoes the photoelectric effect, the electron from that respected K-shell otherwise referred to as the inner shell is ejected via the incident photon. As that electron is excited, vacant space is 'filled' by a neighbouring electron, releasing energy as a photoelectron.</p><p>In short, when a photon has sufficient energy to overcome the electron's binding energy in the K-shell, that atom undergoes the photoelectric effect.</p><p>Each substance owns a unique K-shell binding energy; known as the K-edge. There is a significant spike in attenuation that results just beyond the energy of the K-edge, this peak is unique to every material and holds valuable information about the substance's composition.</p><p>The different photoelectric energies and K-edges are the bread and butter of dual-energy CT. Although most elements in the human body have very low K-edges (0.01-0.53 keV), elements like iodine and calcium have  K-edges of 33.2 keV and 4.0 keV respectively, making them sufficiently larger than surrounding structures and are particularly important in the clinical setting <sup>1-3</sup>.  </p><p>For instance, at 80 kVp a structure that contains no (introduced) iodine, such as the liver, has an attenuation based on its K-edge of x, yet when iodine (33.2 keV) is introduced into that same structure, it has a higher attenuation of y bringing it closer to 80 kVp.</p><p>As 80 kVp is closer to 33.2 keV than 140 kVp, the structures containing iodine will retain less attenuation as the kVp progressed beyond the K-edge of iodine. Therefore, when using two energies, it is possible to delineate structures based solely on their attenuation differences between 80 kVp and 140 kVp.</p><p>A dual x-ray source, tube A (140 kVp) and tube B (80 kVp or 100 kVp) with an angular offset of 90 degrees are preferred offsets for a dual source scanner in the current literature <sup>1-5</sup>.</p><h4>See also </h4><ul><li><a href="/articles/dual-energy-ct-clinical-applications-1">clinical applications of dual energy CT</a></li></ul>

Updates to Link Attributes

Title was removed:
Dual energy CT
Type was removed.
Visible was set to .

Updates to Link Attributes

ADVERTISEMENT: Supporters see fewer/no ads

Updating… Please wait.

 Unable to process the form. Check for errors and try again.

 Thank you for updating your details.