X-ray artifacts

Changed by Mateusz Wilczek, 14 Nov 2023
Disclosures - updated 9 Oct 2023: Nothing to disclose

Updates to Article Attributes

Body was changed:

X-ray artifacts can present in a variety of ways including abnormal shadows noted on a radiograph or degraded image quality, and have been produced by artificial means from hardware failure, operator error and software (post-processing) artifacts. 

There are common and distinct artifacts for film, computed (CR) and direct digital radiography (DR).

Common causes

  • improper handling of the films

  • errors while processing the films

  • patient movement while taking the image

Common artifacts (all forms of radiography)

  • motion artifact

    • due to patient movement resulting in a distorted image

  • image compositing (or twin/double exposure)

    • superimposition of two structures from different locations due to double exposure of same film/plate

  • grid cut-off

  • radiopaque objects on/external to the patient (e.g. jewellery (e.g. necklaces, piercings), clothing (e.g. buttons), hair (e.g. ponytail, hair braids etc.)

  • debris in the housing 

    • debris in the housing caused by the collimator tube can cause small trapezoidal regions, indicative of lead shavings

Film radiography artifacts

  • finger marks

    • improper handling with hands

  • clear film

    • malfunction of the machine or placing the film in the fixer before developer solution

  • static electricity

    • black “lightning” marks resulting from films forcibly unwrapped or excessive flexing of the film

  • crescent-shaped black lines

    • due to fingernail pressure on the film

  • crescent-shaped white lines

    • due to cracked intensifying screen

  • black film

    • complete exposure to light.

  • clear spots

    • air bubbles sticking to film during processing

    • fixer splashed on film prior to developing

    • dirt on the intensifying screen

Computed/digital radiography artifacts

  • detector image lag or ghosting

    • latent image from previous exposure present on current exposure

  • incorrect detector orientation i.e. upside-down cassette

    • spoke like radiopaque lines (case 6)

  • backscatter

    • electronics are visible on the exposed image

    • increased radiation exposure required for portable DR (digital radiography) examinations

  • stitching artifacts

    • occur when two separate DR/CR (digital/computed radiography) images are merged into a single image (see case 3)

  • over exposure

  • dead pixel artifact

  • signal dropout 4

    • large areas of signal loss, due to detector drop

  • speckled radiopaque spots 4

    • due to detector drop

  • detector calibration limitation 4

    • faint radiopaque striping (often vertical) in the background of an image, yet not evident on the anatomy 

    • this artifact should be carefully examined, if it does not interfere with the anatomy, it is not a detector failure/grid cut off, rather a limitation of the detector calibration

    • often seen as lower exposure

  • failure of detector offset correction 4

    • similar to ghosting, however, the digital detector not being calibrated when promoted is the cause 

  • electronic shutter failure 4

    • the digital image often will have obscurely shaped, tight collimation that defies logic

    • often a computer error often fixed with recollimation post exam (this should be explored before re-examination)

  • values of interest misread 4

    • image appears washed out and underexposed

    • this is often due to a largely collimated area of smaller anatomy i.e. a patella protection 

    • tighter digital collimation in conjunction with reprocessing will correctly assign the correct values of interest

  • mid grey clipping 4

    • loss of contrast in areas of different pixel density yet not change in density can be seen i.e. the metal on a knee replacement 

    • due to poor contrast enhancement

  • grid-line suppression failure 4​​

    • faint grid lines present on an image, with no grid cut off

  • -<p><strong>X-ray artifacts</strong> can present in a variety of ways including abnormal shadows noted on a radiograph or degraded image quality, and have been produced by artificial means from hardware failure, operator error and software (post-processing) artifacts. </p><p>There are common and distinct <a href="/articles/radiological-image-artifact">artifacts</a> for <a href="/articles/x-ray-film">film</a>, <a href="/articles/computed-radiography">computed (CR)</a> and <a href="/articles/digital-radiography">digital radiography (DR)</a>. </p><h4>Common causes</h4><ul>
  • -<li>improper handling of the films</li>
  • -<li>errors while processing the films</li>
  • -<li>patient movement while taking the image</li>
  • -</ul><h4>Common artifacts (all forms of radiography)</h4><ul>
  • -<li>motion artifact<ul><li>due to patient movement resulting in a distorted image</li></ul>
  • -</li>
  • -<li>image compositing (or twin/double exposure)<ul><li>superimposition of two structures from different locations due to double exposure of same film/plate</li></ul>
  • -</li>
  • -<li><a href="/articles/grid-cutoff">grid cut-off</a></li>
  • -<li>radiopaque objects on/external to the patient (e.g. <a href="/articles/jewellery-artifact">jewellery</a> (e.g. necklaces, piercings), <a href="/articles/clothing-artifact">clothing</a> (e.g. buttons), <a href="/articles/hair-artifact">hair</a> (e.g. ponytail, hair braids etc.)</li>
  • -<li>debris in the housing <ul><li>debris in the housing caused by the collimator tube can cause small trapezoidal regions, indicative of lead shavings</li></ul>
  • -</li>
  • -</ul><h4>Film radiography artifacts</h4><ul>
  • -<li>finger marks<ul><li>improper handling with hands</li></ul>
  • -</li>
  • -<li>clear film<ul><li>malfunction of the machine or placing the film in the fixer before developer solution</li></ul>
  • -</li>
  • -<li>static electricity<ul><li>black “lightning” marks resulting from films forcibly unwrapped or excessive flexing of the film</li></ul>
  • -</li>
  • -<li>crescent-shaped black lines<ul><li>due to fingernail pressure on the film</li></ul>
  • -</li>
  • -<li>crescent-shaped white lines<ul><li>due to cracked intensifying screen</li></ul>
  • -</li>
  • -<li>black film<ul><li>complete exposure to light.</li></ul>
  • -</li>
  • -<li>clear spots<ul>
  • -<li>air bubbles sticking to film during processing</li>
  • -<li>fixer splashed on film prior to developing</li>
  • -<li>dirt on the intensifying screen</li>
  • -</ul>
  • -</li>
  • -</ul><h4>Computed/digital radiography artifacts</h4><ul>
  • -<li>detector image lag or ghosting<ul><li>latent image from previous exposure present on current exposure</li></ul>
  • -</li>
  • -<li>incorrect detector orientation i.e. upside-down cassette<ul><li>spoke like radiopaque lines (case 6)</li></ul>
  • -</li>
  • -<li>backscatter<ul>
  • -<li>electronics are visible on the exposed image</li>
  • -<li>increased radiation exposure required for portable DR (digital radiography) examinations</li>
  • -</ul>
  • -</li>
  • -<li>stitching artifacts<ul><li>occur when two separate DR/CR (digital/computed radiography) images are merged into a single image (see case 3)</li></ul>
  • -</li>
  • -<li><a href="/articles/over-exposure">over exposure</a></li>
  • -<li><a href="/articles/dead-pixel-artifact">dead pixel artifact</a></li>
  • -<li>signal dropout <sup>4 </sup><ul><li>large areas of signal loss, due to detector drop</li></ul>
  • -</li>
  • -<li>speckled radiopaque spots <sup>4 </sup><ul><li>due to detector drop</li></ul>
  • -</li>
  • -<li>detector calibration limitation <sup>4 </sup><ul>
  • -<li>faint radiopaque striping (often vertical) in the background of an image, yet not evident on the anatomy </li>
  • -<li>this artifact should be carefully examined, if it does not interfere with the anatomy, it is not a detector failure/grid cut off, rather a limitation of the detector calibration</li>
  • -<li>often seen as lower exposure</li>
  • -</ul>
  • -</li>
  • -<li>failure of detector offset correction <sup>4 </sup><ul><li>similar to ghosting, however, the digital detector not being calibrated when promoted is the cause </li></ul>
  • -</li>
  • -<li>electronic shutter failure <sup>4 </sup><ul>
  • -<li>the digital image often will have obscurely shaped, tight collimation that defies logic</li>
  • -<li>often a computer error often fixed with recollimation post exam (this should be explored before re-examination)</li>
  • -</ul>
  • -</li>
  • -<li>values of interest misread <sup>4 </sup><ul>
  • -<li>image appears washed out and underexposed</li>
  • -<li>this is often due to a largely collimated area of smaller anatomy i.e. a patella protection </li>
  • -<li>tighter digital collimation in conjunction with reprocessing will correctly assign the correct values of interest</li>
  • -</ul>
  • -</li>
  • -<li>mid grey clipping <sup>4</sup><ul>
  • -<li>loss of contrast in areas of different pixel density yet not change in density can be seen i.e. the metal on a knee replacement </li>
  • -<li>due to poor contrast enhancement</li>
  • -</ul>
  • -</li>
  • -<li>grid-line suppression failure <sup>4​​</sup><ul><li>faint grid lines present on an image, with no grid cut off</li></ul>
  • -</li>
  • +<p><strong>X-ray artifacts</strong> can present in a variety of ways including abnormal shadows noted on a radiograph or degraded image quality, and have been produced by artificial means from hardware failure, operator error and software (post-processing) artifacts. </p><p>There are common and distinct <a href="/articles/radiological-image-artifact">artifacts</a> for <a href="/articles/x-ray-film">film</a>, <a href="/articles/computed-radiography" title="Computed radiography (CR)">computed (CR)</a> and <a href="/articles/direct-digital-radiography" title="Direct digital radiography">direct digital radiography (DR)</a>.</p><h4>Common causes</h4><ul>
  • +<li><p>improper handling of the films</p></li>
  • +<li><p>errors while processing the films</p></li>
  • +<li><p>patient movement while taking the image</p></li>
  • +</ul><h4>Common artifacts (all forms of radiography)</h4><ul>
  • +<li>
  • +<p>motion artifact</p>
  • +<ul><li><p>due to patient movement resulting in a distorted image</p></li></ul>
  • +</li>
  • +<li>
  • +<p>image compositing (or twin/double exposure)</p>
  • +<ul><li><p>superimposition of two structures from different locations due to double exposure of same film/plate</p></li></ul>
  • +</li>
  • +<li><p><a href="/articles/grid-cutoff">grid cut-off</a></p></li>
  • +<li><p>radiopaque objects on/external to the patient (e.g. <a href="/articles/jewellery-artifact">jewellery</a> (e.g. necklaces, piercings), <a href="/articles/clothing-artifact">clothing</a> (e.g. buttons), <a href="/articles/hair-artifact">hair</a> (e.g. ponytail, hair braids etc.)</p></li>
  • +<li>
  • +<p>debris in the housing </p>
  • +<ul><li><p>debris in the housing caused by the collimator tube can cause small trapezoidal regions, indicative of lead shavings</p></li></ul>
  • +</li>
  • +</ul><h4>Film radiography artifacts</h4><ul>
  • +<li>
  • +<p>finger marks</p>
  • +<ul><li><p>improper handling with hands</p></li></ul>
  • +</li>
  • +<li>
  • +<p>clear film</p>
  • +<ul><li><p>malfunction of the machine or placing the film in the fixer before developer solution</p></li></ul>
  • +</li>
  • +<li>
  • +<p>static electricity</p>
  • +<ul><li><p>black “lightning” marks resulting from films forcibly unwrapped or excessive flexing of the film</p></li></ul>
  • +</li>
  • +<li>
  • +<p>crescent-shaped black lines</p>
  • +<ul><li><p>due to fingernail pressure on the film</p></li></ul>
  • +</li>
  • +<li>
  • +<p>crescent-shaped white lines</p>
  • +<ul><li><p>due to cracked intensifying screen</p></li></ul>
  • +</li>
  • +<li>
  • +<p>black film</p>
  • +<ul><li><p>complete exposure to light.</p></li></ul>
  • +</li>
  • +<li>
  • +<p>clear spots</p>
  • +<ul>
  • +<li><p>air bubbles sticking to film during processing</p></li>
  • +<li><p>fixer splashed on film prior to developing</p></li>
  • +<li><p>dirt on the intensifying screen</p></li>
  • +</ul>
  • +</li>
  • +</ul><h4>Computed/digital radiography artifacts</h4><ul>
  • +<li>
  • +<p>detector image lag or ghosting</p>
  • +<ul><li><p>latent image from previous exposure present on current exposure</p></li></ul>
  • +</li>
  • +<li>
  • +<p>incorrect detector orientation i.e. upside-down cassette</p>
  • +<ul><li><p>spoke like radiopaque lines (case 6)</p></li></ul>
  • +</li>
  • +<li>
  • +<p>backscatter</p>
  • +<ul>
  • +<li><p>electronics are visible on the exposed image</p></li>
  • +<li><p>increased radiation exposure required for portable DR (digital radiography) examinations</p></li>
  • +</ul>
  • +</li>
  • +<li>
  • +<p>stitching artifacts</p>
  • +<ul><li><p>occur when two separate DR/CR (digital/computed radiography) images are merged into a single image (see case 3)</p></li></ul>
  • +</li>
  • +<li><p><a href="/articles/over-exposure">over exposure</a></p></li>
  • +<li><p><a href="/articles/dead-pixel-artifact">dead pixel artifact</a></p></li>
  • +<li>
  • +<p>signal dropout <sup>4</sup></p>
  • +<ul><li><p>large areas of signal loss, due to detector drop</p></li></ul>
  • +</li>
  • +<li>
  • +<p>speckled radiopaque spots <sup>4</sup></p>
  • +<ul><li><p>due to detector drop</p></li></ul>
  • +</li>
  • +<li>
  • +<p>detector calibration limitation <sup>4</sup></p>
  • +<ul>
  • +<li><p>faint radiopaque striping (often vertical) in the background of an image, yet not evident on the anatomy </p></li>
  • +<li><p>this artifact should be carefully examined, if it does not interfere with the anatomy, it is not a detector failure/grid cut off, rather a limitation of the detector calibration</p></li>
  • +<li><p>often seen as lower exposure</p></li>
  • +</ul>
  • +</li>
  • +<li>
  • +<p>failure of detector offset correction <sup>4</sup></p>
  • +<ul><li><p>similar to ghosting, however, the digital detector not being calibrated when promoted is the cause </p></li></ul>
  • +</li>
  • +<li>
  • +<p>electronic shutter failure <sup>4</sup></p>
  • +<ul>
  • +<li><p>the digital image often will have obscurely shaped, tight collimation that defies logic</p></li>
  • +<li><p>often a computer error often fixed with recollimation post exam (this should be explored before re-examination)</p></li>
  • +</ul>
  • +</li>
  • +<li>
  • +<p>values of interest misread <sup>4</sup></p>
  • +<ul>
  • +<li><p>image appears washed out and underexposed</p></li>
  • +<li><p>this is often due to a largely collimated area of smaller anatomy i.e. a patella protection </p></li>
  • +<li><p>tighter digital collimation in conjunction with reprocessing will correctly assign the correct values of interest</p></li>
  • +</ul>
  • +</li>
  • +<li>
  • +<p>mid grey clipping <sup>4</sup></p>
  • +<ul>
  • +<li><p>loss of contrast in areas of different pixel density yet not change in density can be seen i.e. the metal on a knee replacement </p></li>
  • +<li><p>due to poor contrast enhancement</p></li>
  • +</ul>
  • +</li>
  • +<li>
  • +<p>grid-line suppression failure <sup>4​​</sup></p>
  • +<ul><li><p>faint grid lines present on an image, with no grid cut off</p></li></ul>
  • +</li>

ADVERTISEMENT: Supporters see fewer/no ads

Updating… Please wait.

 Unable to process the form. Check for errors and try again.

 Thank you for updating your details.