Presentation
Fracture of the middle third of the left femur 3 months ago. Fall from own height on the current day.
Patient Data
Age: 20 years
Gender: Male
From the case:
Fracture of metallic osteosynthesis screws
{"current_user":null,"step_through_annotations":true,"access":{"can_edit":false,"can_download":true,"can_toggle_annotations":true,"can_feature":false,"can_examine_pipeline_reports":false,"can_pin":false},"extraPropsURL":"/studies/101937/annotated_viewer_json?lang=us"}
Fracture of the middle diaphysis of the femur, fixed with a metal plate and screws, with a fracture and loosening of the lower screws and consequently deviation of the distal segment of the metal plate.
Case Discussion
Surgical re-approach was performed to correct the fracture.